Clean external surface of handpiece by wiping with a damp toothbrush or paper towel. Do NOT run under water. Wipe dry after cleaning.
Insert bur into chuck. (Disregard steps 3 − 8 if you have a MaintenanceFree or MaintenanceFree Ceramic Turbine)
Spray cleaner into drive air line for 2-3 seconds.
Rotate the bur back and forth between your fingers to loosen any debris inside the handpiece head.
Operate handpiece over a paper towel to purge used lubricant and debris.
Repeat steps 3 - 5 until expelled fluid is completely clear.
Operate handpiece until all cleaner has been expelled.
Wipe external surface with a dry towel to remove expelled fluid.
Sterilize handpiece according to manufacturer’s instructions.
POST-STERILIZATION
Allow sterilized handpiece to cool gradually to room temperature. (Disregard steps 2 − 4 if you have a MaintenanceFree or MaintenanceFree Ceramic Turbine)
Verify that the bur is securely in the chuck.
Apply 2 drops of lubricant into drive air line, or spray lubricant into drive air line for 2-3 seconds.
Operate handpiece for 30 seconds over a paper towel to expel excess lubricant from the handpiece.
Wipe external surfaces with a dry towel.
Handpiece is ready for use.
PUSH BUTTON CHUCK MAINTENANCE
Twice a week
Remove bur from handpiece
Place 2 drops of handpiece oil into the chuck
Insert bur into the handpiece.
Run handpiece over a paper towel at 35 psi to expel excess lubricant. If your push button chuck is stiff, after step 2 hold the chuck open and work the
bur in and out to flush out debris.
EXTERNAL MAINTENANCE
Never use any type of cold sterile, alcohol, or disinfectant on your handpiece.
Do NOT submerge handpiece in running water or ultra sonic cleaner.
To remove debris from the outside of the handpiece, use a damp toothbrush, then use a dry paper towel
to wipe off the outside of the handpiece. Discard the paper towel in a biohazard bin.
Clean toothbrush with a cold sterile wipe or in an ultrasonic cleaner for 1 cycle.
Rinse toothbrush thoroughly in water before repeating the cleaning process.
To sterilize, see Pre-Sterilization instructions (on front page).
Check sterilizer temperature. It should be no higher than 276°F (135°C).
OTHER HELPFUL TIPS
Air pressure should be set to 30 − 35 psi. If running MaintenanceFree turbines, air pressure should be set to 40 − 42 psi.
Always insert bur completely into chucks.
Never depress the push button to stop the bur.
Avoid dropping the handpiece.
Do NOT use cold sterile, alcohol or disinfectant on your handpiece.
Check sterilizer temperature. It should be no higher than 276°F/135°C
Before using handpiece or applying a load, be sure to run it for 10 − 15 seconds.